The Multi-Function Calibrator is useful for the calibration of Digital Multimeters, Clamp Meters, Digital Panel Meters, Analog Panel Meters, LCR Meters & Non-Contact Type Tachometers. It is a variable source with different ranges of DC Voltage, AC Voltage, DC Current, AC Current, Frequency, Resistance, Capacitance, Inductance & RPM (Non-Contact Type). It has been designed to calibrate the maximum ranges of a variety of models of multimeters available in the market. The calibrator gives very stable Output. Stability is suitable enough to calibrate 4½ and 4½ DMM and Clamp-on Meters. By using this source along with a high-resolution standard Multimeters user can calibrate even 5½ DMM. Calibrator has a micro controller unit, which handles all the parameter selection and range selection logic. Micro controller drives to the set of relays as per the logic program. Relays rout the outputs.
| Model | ZMMFC 5.5.3 | |
| Input Power Supply | 230V AC ±5% @ 50Hz. | |
| Input Protection | Calibrator is Protected by I/P OV (Over Voltage) & I/P UV (Under Voltage) | |
| DC Voltage Ranges | 200mV, 2V, 20V, 200V, 1000V | |
| AC Voltage (45Hz to 1000Hz) Ranges | 200mV, 2V, 20V, 200V, 1000V | |
| DC Current Ranges | 2mA, 20mA, 200mA, 2000mA, 20A | |
| AC Current (45Hz to 1000Hz) Ranges | 2mA, 20mA, 200mA, 2000mA, 20A | |
| Frequency Range | 45Hz to 1000Hz | |
| Fixed Resistance | 1Ω, 10Ω, 100Ω, 1kΩ, 10kΩ, 100kΩ, 1MΩ, 10Ω, 100MΩ | |
| Fixed Inductance | 10µH, 100µH, 1mH, 10mH, 100mH | |
| Fixed Capacitance | 1nF, 10nF, 100nF, 1µF, 10µF | |
| RPM Range (Non-Contact Type) | 2700 to 60000 RPM | |
| Accuracy @ 20oC to 28oC, RH <75% | AC Voltage: ±0.15 % of reading ±15 digit | DC Voltage: ±0.1 % of reading ±5 digit |
| AC Current: ±0.15 % of reading ±15 digit | DC Current: ±0.1 % of reading ±5 digit | |
| Frequency: ±0.1 Hz | Resistance: ±0.2 % | |
| Inductance: ±2% | Capacitance: ±2 % | |
| RPM (Non Contact Type): ±0.1 % of Reading ±5 Digit | ||
| Operating Temperature | 15oC to 50oC | |
| Power OFF Storage Temperature | 10oC to 70oC | |
| Required Earth to Neutral | Less than 1 V AC | |
| Drift due to ageing | 50 PPM per year. | |
| Short term stability of the set value | 350 PPM | |
| Display | 5½ Digit 7 segment red LED display for Voltage & Current | |
| 4 Digit 7 segment red LED display for Frequency | ||
| Selection of AC/DC Parameters | By up and down keys and bar of LED to indicate selected parameter | |
| Selection of Ranges | By up and down keys and bar of LED to indicate selected range | |
| Setting of Voltage & Current | By decade switch | |
| Setting of Frequency | By multi turn potentiometer | |
| Frequency Range Selection | By Toggle Switch for Normal Mode: 45 to 1 kHz & Filter Mode: 50 to 60 Hz provided at front side. | |
| Resistor mode & Inductance / Capacitance mode Selection | By Toggle switch | |
| R/L/C Range Selection | By Rotary switch | |
| Dimension & Weight | 600mm(W) X 600mm(D) X 170mm(H) | 33kg. (Approx.) |